Peak Areas Not Repeatable

Agilent 7890 Series Troubleshooting
Check the ALS syringe operation. (See the Troubleshooting
section of the sampler user manual:
Replace
the syringe.
• Check for leaks in the inlet, liner (as applicable), and column
connection. (See
"Checking for
• Check sample level in vials.
• Run replicates of known standards to verify the problem.
• Consider if this is the first run. (Has the GC stabilized?)
For a multimode or split/splitless inlet in split mode, also check
for:
• An abnormal split vent restriction. See:
To Run the Inlet Trap Check
To Perform a SS Split Vent Restriction
• To troubleshoot small leaks in a multimode or split/splitless
inlet, see
To Check for Leaks in a Split/Splitless Inlet
Check for Leaks in a Multimode