Blinking Not Ready Light: Detector Hardware Fault/TCD Filament Voltage

Agilent 7890 Series Troubleshooting
If the TCD filament is open, the Not Ready LED blinks and the
GC will not become Ready.
Press [Status]. If the display reads Front detector hardware fault
1
or (Back or Aux detector hardware fault), the problem may be
the TCD filament.
Press [Front Det] (or [Back Det] or [MS/Aux Det]).
2
3
For GCs with firmware B.02.02 or higher, the display will
read either TCD Open Filament for an open filament, or TCD
Bridge Voltage for a shorted filament.
For GCs with earlier firmware: If the display reads Fault: TCD
Filament Voltage, scroll to the Filament setpoint. If the setpoint
is flashing Off, the problem is an open TCD filament.
4
Power cycle the GC.
Press [Front Det] (or [Back Det] or [MS/Aux Det]).
5
Scroll to the Filament and press [Off/No] to disable the TCD
6
filament. Contact Agilent for service.
To continue to use the GC without the TCD:
1
Turn off all TCD parameters. Save the method.
2
Reconfigure the GC to use another detector. Move the column
and set other parameters as needed.
ALS and Detector Symptoms