To Adjust the FID Lit Offset

194

To adjust the FID Lit offset:

Press [Config].
1
Scroll to Front detector or Back detector (wherever the detector
2
is installed) and press [Enter].
Scroll to Lit offset. With the Lit offset line highlighted, enter the
3
new parameter for the detector and press [Enter].
4
Lit offset should be < 2.0 pA or lower than the normal FID
output when lit.