Troubleshooting; If Pcr Yield Is Low When Running Small Reaction Volumes With The Standard Ramp Rate; If No Actual Temperature Is Displayed During The Run; If The Program Does Not Start - Agilent Technologies SureCycler 8800 Setup And User Manual

Table of Contents

Advertisement

4
PCR Programs

Troubleshooting

Troubleshooting

This section describes potential problems and recommended actions.
If PCR yield is low when running small reaction volumes with the
Standard ramp rate

If no actual temperature is displayed during the run

The thermal block unit may not be correctly seated on the instrument.

If the program does not start

The thermal block unit may not be correctly seated on the instrument.
58
If you are using sealing films/mats, shorten the duration of the
denaturation step during cycling. Denaturation times as short as 3–5
seconds may be appropriate. Alternatively, switch to using capped
reaction tubes.
Increase the reaction volume to 50 μl.
Try switching to the Fast ramp rate.
Check that the unit is installed correctly and reboot the cycler.
Check that the unit is installed correctly and reboot the cycler.
Agilent SureCycler 8800 Setup and User's Guide