4.3.2 Measuring range

sequence of 0.125 ms and frequency of 4 kHz, the error is 0.4 dB.
Linearity Range: The measurement result can be considered correct only if it re-
mains within the linearity range. Otherwise the measurement result error exceed the
tolerance limit. Sometimes the linearity range is called the "measuring range".
Dynamic Range: The dynamic range is a range between the own noise and the
maximum input signal level. The dynamic range is the maximum range that the
sound level meter can display. A result close to the own noise cannot be treated as a
linear result.
Peak C Range: The Peak C Range is a linear Peak C measuring range. The Peak
C measurement within this range can be considered to be correct.

4.3.3 ICCP Power

4.3.4 Profiles 1– 3
keys, you can choose the following options: A, B, C and Z (Z correction) means
"no correction". It is sometimes called the "fixed correction" or "linear correction".
Detector:
You must define the filter for Profiles 1–3. With the <◄> and <►> keys, you can
choose the following options: Fast, Slow and Pulse.
Mode:
It defines the integration mode for Profiles 1–3. With the <◄> and <►> keys,
you can choose the following options: SPL, PEAK, LEQ, MAX and MIN.
SWN Save:
With this option, you can define the data to be saved as an SWN file, since their
source is the SWN file for Profiles 1–3. This option is not represented on the
SW-BA-e-1710
The MEAS.Range includes the Linearity Range, Dynamic
Range and Peak C Range.
With the new algorithm, there is only one measuring range, so
switching between the ranges is not required. The algorithm
meets the requirements of a pulse frequency response of up to
0.25 ms with an error being as low as 0.1 dB at 4 kHz. At a pulse
In the ICCP menu, you can adjust the power supply for all ICCP
sensors using a DC power source of 24 V/4 mA. The ICCP power
should be deactivated before connecting another sensor or before
a direct connection to the signal source. Press the navigation keys
<◄> and <►> to choose an option.
The Profiles 1–3 menu allows you to define the filter, detector,
mode and options for SWN file storage. You can choose the op-
tions by pressing the navigation keys <▲> and <▼>.
Filter:
You must define the filter for Profiles 1–3. With the <◄> and <►>
25