4. Operation and setting the menu

Menu tree

4.1 Function

4.2 Calibration

calibration using the sound calibrator. Calibration By Calibration Factor allows the
user to manually adapt the calibration factor.
20
Press <MENU> to show the next menu. All the measurement pa-
rameters can be set in the menu.
Select Function and press <ENTER> to open this menu. You can
choose 2 types of measurement. Level measurement and
1/1-octave measurement. Press the navigation keys <▲> and
<▼> to choose the measurement mode. Press <ENTER> to save
the setting and return to the previous menu. Press <ESC> to re-
turn to the previous menu.
Select Calibration and press <ENTER> to open this menu.
The microphone sensitivity is influenced by many factors, such as
temperature, humidity or air pressure. This is why you must per-
form at least one calibration before the measurement.
Two calibration methods are available. By Measurement and By
Calibration Factor. Choose By Measurement to perform the