Measuring Principle - sauter TU-US Instruction Manual

Digital ultrasonic thickness gauge
Hide thumbs Also See for TU-US:
Table of Contents

Advertisement

Instruction Manual
-
Four transducer models are available for special
applications included coarse grain material and
high temperature applications.
-
Zero adjustment function
-
Sound velocity calibration function
-
Two- point calibration function
-
Two measurement modes: Single point mode
-
Coupling status indicator showing the coupling
status
-
Battery indication indicates the rest capacity of
the battery
-
"Auto sleep" and "Auto power off" function to
conserve battery's life
Optional software to transfer the memory data to
PC
-
Optional thermal mini- printer to print the
measured data via RS-232 port.

1.3 Measuring principle

The digital ultrasonic thickness gauge determines the
thickness of a part or a structure by accurately measuring
The time required for a short ultrasonic pulse generated by
a transducer to travel through the thickness of the material,
to reflect from the back or inside surface and be returned
to the transducer. The measured two-way transit time is
devided by two to account for the down-and-back travel
path, and then multiplied by the velocity of sound in the
material. The result is expressed in following relationship:
v
=
H
Where: H ----
thickness of the test piece
v ----
sound velocity in the material
t ----
the measured round-trip transit time
1.4 Configuration
No.
Item
Stan-
1
Main body
dard
2
Transducer
Con-
figu-
3
Couplant
ration
4
Instrument Case
5
Operating
Manual
6
Screwdriver
7
Alkaline battery
8
Optio-
9
Transducer: N02
nal
10
Transducer: N07
Con-
11
Transducer: HT5
figu-
12
Mini thermal
ration
printer
13
Print cable
14
DataPro for
Thickness
Gauge
TU_US-BA-e-1110
Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: info@sauter.eu
TU_US
Scan mode
×
t
2
Quan-
Note
tity
1
1
Model:
N05/90°
1
1
1
1
2
AA size
See
Table3-1
1
1
1
For use
on the
PC
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
15
Communication
Cable
1.5 Operation conditions
Temperature:
-20°C up to +60°C
Storage temperature: -30°C up to 70°C
Relative humidity:
≤ 90%
In the surrounding environment any kind of vibrations
should be avoided, as well as magnetic fields, corrosive
medium and heavy dust.
2.
Structure feature
2.1 Instrument appearance
1 Main body
2 Transducer
MT200
1
MiTec h
2.2 Parts of the main body
7
6
MT200
2
5
8
MiTec h
3
4
1 Communication Socket
2 Aluminium case
3 Belt hole
4 Battery cover
5 Keypad
6 LCD Display
7 Socket of transducer (no polarity)
8 Control plate (inbuilt)
9 Aluminium case
10 Label
1
2
10
9
THICKNESS GAUGE
POWER: 2 X 1.5V
SN:
OPERATION GUIDE
1. Plug in the transducer
2.
Power On/Off
3.
Backlight On/Off
4.
Probe Zero
5.
Switch Selection
6.
Save/Delete
7.
Exit
8.
Enter
MiTech Inc. Ltd
2

Advertisement

Table of Contents
loading

Table of Contents