Calibration; Firmware Updating; Installation Of Usb Driver - sauter SW 1000 Instruction Manual

Sound level
Hide thumbs Also See for SW 1000:
Table of Contents

Advertisement

sound source was clearly
changed.
A key does not work.
Slow response during op-
eration.
Measurement data cannot
be saved.
The printer does not print
the measurement data.

6.3 Calibration

The sound level meter is factory calibrated. Perform regular calibration to maintain
the measuring accuracy. BSWA offers calibration services for acoustic products.

6.4 Firmware updating

The firmware of SW 1000/SW 2000 can be updated using the USB interface. The
following equipment components must be available:
 the sound level meter SW 1000/SW 2000 must be deactivated (identifier HWID
P0274 or higher),
 miniUSB cable (included in delivery),
 external power supply (included in delivery),
 updating firmware (to be downloaded from the BSWA website),
 USB driver (CP210x driver by Silicon Labs).

6.4.1 Installation of USB driver

the driver is installed.
44
 Weak contact between the microphone and casing:
send the casing to the manufacturer.
Damaged key: send the device to the manufacturer.
Too many files in the microSD card: delete the dam-
aged files.
 Check the logger settings.
 Format the memory card with FAT32.
 Replace the microSD card for a new memory card
with a maximum capacity of 4 GB.
 Check the printer settings.
 Check if paper is properly placed in the printer.
Unpack and install the driver according to the fol-
lowing procedure. Select the X86 version for a
32-bit operating system or the X64 version for a
64-bit operating system.
☆ Hint: Do not connect the sound level meter to
the computer during installation.
Following the installation instructions, confirm the
Licence Agreement and then click Continue until