10.13 Bluetooth

For products equipped with Bluetooth, you can choose whether you want to
communicate with the PC software via Bluetooth.
When Bluetooth is switched on, the Bluetooth symbol is displayed in the status
bar. If the client programme is installed on the PC, switch Bluetooth on in the
"System settings" of the device and connect the computer to Bluetooth. After
successful synchronisation, the software will use Bluetooth connection mode to
establish a connection and a message will appear in the bottom right corner of
the software indicating that the connection via Bluetooth is successful.

10.14 Simple mode

After the simple mode has been switched on, go back to the measurement page
for the standard measurement. After the standard measurement, the device
automatically switches to sample measurement mode.

10.15 Measuring mode

SCI includes the specular reflection measurement mode, SCE excludes the
specular reflection measurement mode. In this device, the SCI/SCE test mode
is switched by the traditional way of setting mechanical optical traps. In SCI
measurement mode, the motor drives the paddles to block the mechanical
optical traps, and in SCE measurement mode, the paddles are opened. In
standard measurement, the device automatically performs SCI and SCE
measurement, and the test time is about 3 seconds. When measuring the
sample, the device measures according to the measurement mode set by the
customer. The customer can set the measurement mode as SCI, SCE or I+E
according to the needs of the measurement products, and some models only
have some options. I+E is the SCI+SCE mode. The measurement time of
SCI/SCE alone is about 1.5 seconds, and the simultaneous measurement of
SCI+SCE takes 3.2 seconds. When the current measurement mode of the
device is SCI (SCI is displayed in the working status area), the device only
checks the SCI data of the sample; when the display mode is set to SCE, the
corresponding chromaticity data is displayed as "-", and the spectral data and
colour index are not displayed.

10.16 Display Setting

Click on Display Settings in the
main menu. Here you can set
whether the colour deviation and
the display of the test results should
be activated.
If the colour deviation is switched on,
the colour deviation of the sample
JCS-BA-e-2410
en | 25