Download Print this page

Calibration (Adjustment); Measuring Procedure - sauter TB-US Instruction Manual

Ultrasonic material thickness gauge

Advertisement

6. Calibration (Adjustment)

1. Put a little oil on the zero plate 4-12.
2. The calibration button 4-8 is pressed and 'CAL' appears in the display. CAL' is
the abbreviation for calibration.
3. The
sensor
The display "Measuring process in progress" ((•)) is shown when the measuring
process has been successfully established by the transmit & receive process.
The display shows '5.0' mm (target thickness of the zero plate) and 'CAL'
alternately. Once the display value has stabilized, press the 'CAL' key 4-8 to
confirm. The instrument then switches back to the measuring mode.
4. The calibration (adjustment) is now completed and is automatically stored in the
device.

7. Measuring procedure

1. The power button 4-4 is pressed to switch on the unit.
2. The mm/inch key 4-3 is pressed to select the correct measuring unit.
3. The measuring head 4-7 is placed on the material surface to be measured,
provided that the material code has been selected correctly. Make sure that the
"Measuring process in progress" indicator is visible and the character ((-)) 4-11
is displayed. The display shows the measurement result.
4. The measurement result is retained until a new measured value is measured.
The last value remains on the display until the instrument is switched off.
5. The device can be switched off either by pressing the on/off key or by the auto
power off function, one minute after the last key has been pressed.
6
4-7
is
pressed
onto
the
zero
TB_US-BA-e-2020
plate.

Advertisement

loading

This manual is also suitable for:

Tb-us-red