GE Medical Systems LOGIQ ?200 Service Manual page 65

Ultrasound scanner
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
3–6–2
LED List
The table below shows the LED list for LOGIQ α200.
LOCATION
NAME
MST ASSY
V5
WOTOUT
D3~D10
DSC ASSY
DS1
DS2
DS3
DS4
DS5
HV ASSY
DS1
SMPS ASSY
GREEN
LED
Three LED (DS2~DS4) should be blinked when the system operated with the Probe.
3–6–3
DIP Switch
The table below shows the DIP Switch list for LOGIQ α200.
LOCATION
NAME
MST ASSY
S1
DSC ASSY
S1
3–6–4
Reset Switch
Reset switch (S2) on edge of MST Assy used for resetting the system.
TABLE 3–7
DESCRIPTION
+5V for digital
Watchdoc timer out, Reserved
Reserved
+5V for digital
Mapping clock
Mapping clock
Mapping clock
DSP running
HV ASSY status (ON:normal)
SMPS ASSY status (ON:nor-
mal)
Note
TABLE 3–8
DIP SWITCH Setting
Switch No.
1
Initiate the SRAM data to factory
7
(ON:Other countries, OFF:Japan)
2~ 6
8
Select NTSC or PAL (ON:PAL, OFF:NTSC)
1 ~8
3–13
LOGIQ α200 SERVICE MANUAL
LED
POSITION
Edge of board
Edge of board
Edge of board
Edge of board
Edge of board
Edge of board
Edge of board
Edge of board
Edge of board
SMPS ASSY
DESCRIPTION
Setup value
Select AAF Filter
Not Assigned
Not Assigned
2138853
NORMAL
ABNORMAL
ON
OFF
ON
OFF
Refer to
NO
Note
Note
ON
OFF
ON
OFF
ON
OFF
POSITION
Edge of board
Edge of board
SYSTEM CONFIGURATION