GE MEDICAL SYSTEMS
REV 5
6–4–2
Replacement Procedure (Continued)
(D) Final Procedure
1.
Connect four cables with the appropriate connectors. For the details of cables, refer to 6–4–2, (B),
Disconnecting Cables.
2.
Re–assemble the removal parts.
3.
Plug the system and power it ON.
4.
Perform all SMPS Assy checks and leakage current test.
5.
Perform functional checks to verify that the system is functioning properly. Refer to Chapter 4, FUNCTIONAL
CHECKS, in this manual.
6.
This completes the replacement of the SMPS Assy Unit.
LOGIQ α200 SERVICE MANUAL
6–109
2138853
RENEWAL PARTS