6–2–39 AC Fan Assy (FRU No. 417)
Time Required
45 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–45.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
4.
Remove the Top Cover (FRU 409). Refer to 6–2–32.
5.
Disconnect one connector (CN01).
6.
Unscrew four screws (1 – 4) on the fan.
7.
Pull out the Fan Assy outwards.
LOGIQ α200 SERVICE MANUAL