Probe Leakage Current Test (Continued)
Data Sheet for Transducer Source Leakage Current
Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than
the value shown in the table below (TABLE 7– 4).
1. Type BF Applied Part Leakage Current Limits
[Non–conductive (floating) Surface and Cavity Transducers]
COUNTRY
USA
OTHERS
2. Type CF Applied Part Leakage Current Limits
[Surgical Transducer and ECG Connections]
COUNTRY
USA
OTHERS
Transducer tested:
LOGIQ a200
Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
TABLE 7– 4
MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT
NORMAL
OPEN
CONDITION
GROUND
0.05mA
0.05mA
0.1mA
0.5mA
NORMAL
OPEN
CONDITION
GROUND
0.01mA
0.01mA
0.01mA
0.05mA
TABLE 7– 5
DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT
Tester Power
Polarity Switch
NORM
OPEN
NORM
CLOSED
REV
OPEN
REV
CLOSED
NORM
OPEN
NORM
CLOSED
REV
OPEN
REV
CLOSED
7–13
LOGIQ a200 SERVICE MANUAL
REVERSE
POLARITY
0.05mA
0.5mA
REVERSE
POLARITY
0.01mA
0.05mA
Tester Ground
or Neutral Switch
2138853
OPEN
NEUTRAL
0.05mA
0.5mA
OPEN
NEUTRAL
0.01mA
0.05mA