Instrument Accuracy (Average Path) Test; Specification; Description - Keysight Technologies E4417A Service Manual

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

2
Performance Tests

Instrument Accuracy (Average Path) Test

Specification

Electrical Characteristics
Accuracy
[a] This performance limit does not include the corresponding sensor power linearity specification.

Description

The power meter accuracy is verified for various power inputs. There are two
different methods described in this section that can be used to perform this test.
One of these methods has a higher measurement uncertainty value associated
with it.
Test Procedure 1
The measurement uncertainty of this test procedure is relatively high at power
NOTE
levels of 30 μW and below, as it uses a standard 11683A Range Calibrator. As
such, this test procedure should be used for indication only.
Equipment
– Range calibrator: Keysight 11683A.
– Power sensor cable: Keysight E9288A.
38
Performance Limits (maximum)
[a]
±0.5% or 0.02 dB
Keysight E4416A/E4417A Power Meters Service Guide