Procedure

1 Ensure that the power meter has been powered on for at least 30 minutes
2 Connect the equipment as shown in
3 Turn the power reference on by pressing:
4 Observe the reading on the frequency counter.
5 Remove the power meter cover.
6 Adjust A2L9 to obtain a 50.0 ±0.5 MHz indication on the frequency counter.
Keysight E4416A/E4417A Power Meters Service Guide
before making any measurements.
frequency.
Zero
,
Power Ref On
Cal
Zero
,
,
Cal
More
Cal
If it is 50.0 ±0.5 MHz, no adjustment of the power reference oscillator
frequency is necessary. If it is not within these limits, adjust the power
reference oscillator frequency as described in step 5 and step 6.
Refer to
Figure 3-2
for the position of A2L9.
Figure
3-1. Set up the counter to measure
on the Keysight E4416A.
,
on the Keysight E4417A.
Power Ref On
Adjustments
3
75