Keysight 478A Sensors: Additional Information

The Voltage Reflection Coefficients (VRC) of the Keysight 478A sensor must be
known before attempting to perform the Output Standing Wave Ratio (SWR) Test.
The VRC values required for this test are as follows:
– Frequency 50 MHz, with the mount resistance of the Keysight 432A set to 200
– Frequency 50 MHz, with the mount resistance of the Keysight 432A set to 100
For reference only, typical values of these VRC values are as follows:
Using incorrect VRC values in the Output SWR calculations will introduce
NOTE
significant errors.
Keysight E4416A/E4417A Power Meters Service Guide
Keysight 432A Power Meters & 478A Sensors: Additional Information
(recorded as
Γ
).
1
(recorded as
).
Γ
2
0.0012
Γ 1
Γ 2
0.3317