28 Verify that the output SWR is not greater than the maximum value shown in
The
NOTE
network analyzer. For clarification, above
is measured at 100 Ω.
Table 2-8
Output SWR Result
[a] For instruments with the 50 MHz 1 mW output (power ref) connector mounted on the front panel.
[b] For instruments with the 50 MHz 1 mW output (power ref) connector mounted on the rear panel. This applies to models E4416A and E4417A
with Option 003.
Keysight E4416A/E4417A Power Meters Service Guide
the following table:
Γ 1
Γ 2
and
values are measuredfor the thermistor mount (478A) using the
Max
[a]
1.06
[b]
1.08
Performance Tests
Γ 1
is measured at 200 ohm while
Actual Resul t
__________________
__________________
2
Γ 2
67