Procedure; Keysight E932X Power Sensor (Average Path) Functional Test; Description; Equipment - Keysight Technologies E4417A Service Manual

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

Keysight E932x Power Sensor (Average Path) Functional Test

Description

This test verifies that the meter/sensor combination can make average path RF
measurements in both upper and lower power ranges.
Keysight E-Series power sensors operate over a wide dynamic range: the value is
sensor dependent and ranges from −70 dBm to +44 dBm. However as there is an
amplifier in these power sensors, the voltages presented to the power sensor
connector are always within the range of voltages available from the Keysight
11683A range calibrator.

Equipment

– Power sensor cable: Keysight E9288A.
– Reference attenuator: Keysight 11708A.
– Power sensor: Keysight E9321A.

Test Setup

Refer to

Procedure

Use the following procedure to perform the test:
The procedure details the key presses required on the Keysight E4416A. For the
NOTE
Keysight E4417A the equivalent key presses must be performed on both
channels.
Keysight E4416A/E4417A Power Meters Service Guide
Figure 2-7
and
Figure 2-8
below.
Performance Tests
2
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4416a

Table of Contents