Figure 2-8 Sensor (Average Path) Functional Test Setup With - Keysight Technologies E4417A Service Manual

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

4 Zero the instrument:
5 Press
6 Connect the 30 dB attenuator as shown in
Figure 2-8
7 Turn on the power reference:
8 Switch to the power sensor's lower range:
9 Verify that the display reads −30 dBm ±1 dB. Record the reading in
Keysight E4416A/E4417A Power Meters Service Guide
Zero
Press
,
Zero
. Wait approximately 25 seconds for the
Cal
wait symbol to disappear.
Cal
to calibrate the power meter.
Power Meter
CHANNEL A
Sensor (Average Path) Functional Test Setup With Attenuator
Zero
Press
,
Power Ref On
Cal
– Press
Channel
.
– Scroll through the Range values by pressing
LOWER.
– A pop-up window is displayed.
– Press
Done
.
Figure
Power Sensor
30 dB attenuator
Power sensor cable
Performance Tests
2-8.
Change
. Select a value of
Table
2
2-6.
55