Keysight Technologies E4417A Service Manual page 39

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

2
Performance Tests
There is a small risk of mild electrical shock from the chassis of the 11683A
WARNING
Option H01 range calibrator unless the power meter is connected.
3
4 Switch the power meter on.
5 Preset the instrument:
6 Set the display units to watts:
7 Set the display resolution:
8 Set the range calibrator as follows:
When switching the range calibrator to STANDBY, allow enough time for the
NOTE
range calibrator to settle to its zero value before attempting to zero the power
meter. This settling would appear on the power meter display as downward drift.
When the drift has reached minimum, (typically less than 60 seconds), the range
calibrator is settled.
44
Figure 2-5
shows the HI/LO outputs of the DC calibrator connected to the DC
reference input of the range calibrator. Ensure that the Guard and LO output of
the DC calibrator are floating, and are not connected to the earth ground.
Preset
Press
then
Confirm
Local
Meas
Press
,
Units
Display
,
Meas
Press
then press
Display
– RANGE: 100 mW
– POLARITY: NORMAL
– FUNCTION: STANDBY
– REF SWITCH: EXT
– LINE: OFF
.
,
.
W
Resolution
until
Keysight E4416A/E4417A Power Meters Service Guide
4
is selected.