Keysight Technologies E4417A Service Manual page 46

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

11 Round the DVM reading to two decimal places.
12 Disconnect the DVM negative input lead from the VRF connector on the
13 Observe the DVM reading. Record the reading as V
14 Calculate the Power Reference Oscillator power using equation 2-4.
15 Verify that P
Table 2-5
Power Reference Level Test Result
Min
0.991 mW
Keysight E4416A/E4417A Power Meters Service Guide
Record this reading as V
– V
___________
1
Keysight 432A. Reconnect it to the Keysight 432A chassis ground.
– V
___________
comp
is within the limits shown in
meas
in the table.
.
1
Max
1.009 mW
Performance Tests
.
comp
Table 2-5
and record the reading
Actual Resul t
__________________
2
51

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4416a

Table of Contents