Keysight Technologies E4417A Service Manual page 51

Power
Hide thumbs Also See for E4417A:
Table of Contents

Advertisement

2
Performance Tests
10 Switch to the power sensor's upper range:
11 Verify that the display reads −30 dBm ± 1 dB. Record the reading in
12 Turn on the power reference:
13 Switch to the power sensor's lower range:
14 Disconnect the 30 dB attenuator and reconnect the test setup as shown in
15 Turn on the power reference:
16 Verify that an overload error is displayed on the status line at the top of the
56
– Press
Channel
.
– Move to the Range parameter using the up/down arrow keys.
– Scroll through the Range values by pressing
UPPER.
– Press
Done
.
Zero
Press
,
Power Ref Off
Cal
– Press
Channel
.
– Move to the Range parameter using the up/down arrow keys.
– Scroll through the Range values by pressing
LOWER.
– Press
.
Done
Figure
2-7.
Zero
Press
,
Power Ref On
Cal
power meter's display and that the display reads 999.99 dBm.
Keysight E4416A/E4417A Power Meters Service Guide