2
Performance Tests
Table 2-4
Instrument Accuracy Results
DC
Effective
Calibrator
Power
Setting
Applied
458.31 μV
3.1623 μW
1.4493 mV
10 μW
4.5831 mV
31.6228 μW
14.493 mV
100μW
45.831 mV
316.2278 μW 64
144.93 mV
1 mW
459.12 mV
3.1623 mW
1.4588 V
10 mW
4.6872 V
31.6228 mW
15.588 V
100 mW
46
Power
Meas Unc
Lower Limit
Meter
Fil ter
Setting
256
0.028 μW
3.0977 μW
256
0.027 μW
9.904 μW
256
0.027 μW
31.428 μW
64
0.073 μW
99.49 μW
0.104 μW
314.72 μW
16
0.2 μW
0.9954 mW
16
0.9 μW
3.1477 mW
16
1.6 μW
9.954 mW
16
4.8 μW
31.477 mW
16
18.5 μW
99.54 mW
Ch A
Ch B
Resul t
Resul t
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
Keysight E4416A/E4417A Power Meters Service Guide
Upper Limit
GuardBan
d Lower
Limit (in
μW)
3.2269 μW
3.1257
10.096 μW
9.931
31.819 μW
31.455
100.51 μW
99.563
317.74 μW
314.824
1.0047 mW
995.6
3.1769 mW
3148.6
10.046 mW
995.6
31.769 mW
31481.8
100.46 mW
99558.5
GuardBan
d Upper
Limit (in
μW)
3.1989
10.069
31.792
100.437
317.636
1004.5
3176
10044.4
31764.2
100441.5