Configuring The G1369C Lan Card Using Bootp - Agilent Technologies Cary 8454 Installation Manual

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

LAN Communication, Installation, Connection and Configuration
6
Using the Agilent BootP Service

Configuring the G1369C LAN Card Using BootP

In order for an internal G1369C LAN card to use BootP, the default setting:
bootp=YES must be set. For more details please see the user manual of your
G1369C LAN interface card.
To change the setting, the card must be reset; this is done each time the
instrument with the G1369C LAN card inserted is power cycled.
The Agilent BootP Service uses a file called bootptab were all the
configuration information is stored. The BootP Service program on the
Agilent ChemStation CD-ROM allows easy editing of the bootptab file. If a
BootP Service is already installed, the following information must be supplied
to the system administrator to be able to correctly set up the bootptab file.
# CAG Bootptab file
# global defaults
global.defaults:\
sm=255.255.255.0:\
gw=<gateway address - ask your system administrator>
#8454 lab
UV-Vis:\
ht=1:\
ha=0030D30D3277:\
ip=<ip address - ask your system administrator>
sm=255.255.255.0:\
gw=<gateway address - ask your system administrator>
vm=auto:\
hn:\
bs=auto:\
T145=64:T146=01:T147=01
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
89