Agilent Technologies Cary 8454 Installation Manual page 54

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

4
Installing Accessories
Long Path Cell Holder
2 If your sample cell is not one of the standard lengths, position the cell stops
3 Because the light beam is collimated, absolute positioning of the sample cell
4 Install the sample cell between the clamps. Slide it against the cell stop. If
5 Allow the clamps to close against the sample cell, see Figure 18. Tighten the
Figure 18
54
as for a long path sample cell, i.e. move all cell stops in Figure 17 to the left.
is not important; however, consistent cell positioning gives more consistent
results. The cell stops make it easier to consistently position your sample
cells.
your sample cell is not one of the standard lengths, position it so the clamps
are gripping the mid-point of the cell.
four screws that hold the sample cell in place.
Installed 10-cm cylindrical cell in the long path cell holder
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide