Agilent Technologies Cary 8454 Installation Manual page 25

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

1 Your line voltage must be between 100–240 V AC with a line frequency of 50
2 Make sure the power switch of the spectrophotometer is in the off position,
3 Plug the female end of the power cord into the power input socket at the
Installing the Waste Tubing
There is a waste outlet for the sample area. It is used to drain any liquids
which may come from the sample area in case of a leak or a spill to a waste
container. To connect your waste tubing to the spectrophotometer:
1 Lift the front part of the spectrophotometer up so that you can locate the
Figure 3
2 Connect the waste tubing which comes with your spectrophotometer to the
3 Put the other end of the waste tubing into the waste container.
4 Make sure that the waste tubing has a constant down slope from the
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
or 60 Hz. There is no voltage selector on your spectrophotometer because
the power supply has automatic wide ranging capability.
that is, the pushbutton stands out and the LED inside the switch is off, see
Figure 1.
rear of the spectrophotometer. Plug the male end of the power cord into
your electrical outlet.
waste outlet underneath the sample pan.
Connecting the waste tubing
waste outlet underneath the sample pan.
spectrophotometer to the waste container.
Installation and Start Up
Installing your UV-Visible Spectrophotometer
2
25