Installing the Filter Wheel Assembly

Installing the filter wheel assembly is fairly simple. The procedure is the same
for all three cell holders. You will need a Phillips screwdriver to install the
filter wheel assembly. However, if you install the filter wheel on the long path
cell holder, instead of a screwdriver, a 2.5-mm hex wrench is required.
1 Remove the cell holder from the sample compartment.
2 Remove the two screws shown in Figure 21.
3 Position the filter wheel assembly clamp as shown in Figure 21. Replace the
Figure 21
4 Tilt the outer edge of the clamp up slightly.
5 Bring the filter wheel assembly against the cell holder so that the clamp fits
6 Tighten the screws that hold the clamp.
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
two screws but do not tighten them.
Installing the optical filter wheel assembly
in the slot of the filter wheel assembly. The mounting ridges on the filter
wheel assembly should fit into the holes in the clamp. The base of the filter
wheel assembly should be against the base of the cell holder.
Installing Accessories
Optical Filter Wheel
Screw
Clamp
4