6
LAN Communication, Installation, Connection and Configuration
Using the Agilent ChemStation Configuration Editor to Configure the Spectrophotometer

Configuring an Agilent Cary 8454 UV-Visible Spectrophotometer

This section describes how to configure an Agilent Cary 8454 UV-Visible
spectrophotometer for LAN communication.
1 Start the Agilent ChemStation Configuration Editor by double-clicking on
2 Select the Configure > Instruments menu option.
3 Select the Instrument Type UVVis, enter an Instrument Name for your
Figure 33
4 Select the Spectrophotometer 8454 in the Modules tab, select the LAN
5 Select Identify by IP Address and type in the IP address of the LAN
92
the Configuration Editor icon in the UV-Visible ChemStation start-up
group, by choosing Start > All Programs or All Apps > UV-Visible
ChemStations > UV-Vis Configuration Editor.
configuration (e.g. 8454 UV-Visible System [1]), choose an Initial Screen
Window Size, and click OK.
Instrument configuration dialog.
option, and click Add to add to the Selected Modules box.
interface card. (If a DNS server is installed you can also define the host
name.)
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide