Agilent Technologies Cary 8454 Installation Manual page 77

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

3 In the 'Local Area Connection Properties' dialog (see Figure 27) click
If the device is not working properly or no network interface card is configured, please refer
NOTE
to the hardware manual of your network card and the PC for how to install and configure a
network interface card.
4 In the 'This connection uses the following items:' section of this dialog make
If the TCP/IP protocol is not installed, use the Install button, select Protocol from the list of
NOTE
network components to install and press Add.
From the 'Manufacturers:' select 'Microsoft' and from the 'Network Protocol:' select the
'Internet Protocol Version 4(TCP/IPv4)' and press 'OK'. If required, reboot your computer.
5 Select the 'Internet Protocol Version 4(TCP/IPv4)' and click Properties.
If the PC is connected to your companies LAN, do not change any setting. Note down the
NOTE
'Subnet Mask' if available. Ask your responsible IT consultant for this mask if not available
and request a fixed IP address for your instrument, if you want to use the existing LAN for
the instrument connection.
For further details please see"Connecting the Spectrophotometer to the PC".
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
LAN Communication, Installation, Connection and Configuration
Configure... to make sure the device is working properly.
sure the 'Internet Protocol Version 4(TCP/IPv4)' is available and checked.
Installing and Configuring TCP/IP on the PC
6
77

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents