Power Cords; Space - Agilent Technologies Cary 8454 Installation Manual

Uv-visible spectroscopy system
Hide thumbs Also See for Cary 8454:
Table of Contents

Advertisement

2
Installation and Start Up
Site Requirements for Your Spectrophotometer

Power cords

Different power cords are offered as options with the spectrophotometer. The
female end of each of the power cords is identical. This connects to the
power-input socket at the rear of the spectrophotometer. The male end of each
of the power cords is different and designed to match the wall socket of a
particular country or region.
Do not replace the mains power cord with one that is inadequately rated. When replacing
NOTE
the mains power cord, use only a mains power cord with a rating equivalent to the one
supplied with the instrument. See the instrument installation guide for detailed mains
power cord rating information.
Always operate your instrument from a power outlet which has a ground
WARNING
connection. Always use the power cord designed for your region.
Do not position the equipment so that it is difficult to operate the
disconnecting device.

Space

The spectrophotometer is 34.4 cm (13.5 inches) wide, 56.0 cm (22.0 inches)
deep and 18.5 cm (7.3 inches) high (see Figure 1 on page 22). It weighs 14 kg
(30.86 lbs) and will fit on almost any desk or laboratory bench. It needs an
additional 2.5 cm (1.0 inches) of space on either side to allow for the
circulation of air for cooling, and approximately 8 cm (3.1 inches) in the rear
for electrical connections.
The computer, including the keyboard, is approximately 50 cm (19.7 inches)
wide and 60 cm (23.6 inches) deep.
Additional space is required for accessories like pumps and autosamplers.
18
Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide