Recommended Separation Distances - GE Solar 8000M/i Service Manual

Patient monitor
Table of Contents

Advertisement

Appendix B – Electromagnetic compatibility: Electromagnetic compatibility (EMC)

Recommended separation distances

Rated maximum output
power of transmitter in
Watts
0.01
0.1
1
10
100
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2026265-075C
The table below provides the recommended separation distances (in
meters) between portable and mobile RF communications equipment and
the Solar 8000i.
The Solar 8000M/i is intended for use in the electromagnetic
environment on which radiated RF disturbances are controlled. The
customer or the user of the Solar 8000M/i can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the Solar 8000M/i as recommended below, according
to the maximum output power of the communications equipment.
Separation distance in meters (m) according to frequency of transmitter
a
150 kHz to 80 MHz
P
d = 1.2
0.12
0.38
1.2
3.8
12
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated
using the equitation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all instances. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Solar 8000M/i patient monitor
a
80 MHz to 800 MHz
P
d = 1.2
0.12
0.38
1.2
3.8
12
a
800 MHz to 2.5 GHz
P
d = 2.3
0.23
0.73
2.3
7.3
23