GE Solar 8000M/i Service Manual page 166

Patient monitor
Table of Contents

Advertisement

Appendix A – Technical specifications: Tram modules and Solar parameter functionality
Analog output:
Alarm limit range:
Messages sent to host:
GE:
Pulse search, Poor signal quality detected, Probe is off patient, Probe or module malfunction
Masimo: Probe or module malfunction, Artifact detected, Probe is off patient, Low quality, Pulse search
Nellcor:
Probe is off patient, Low quality, Pulse search
Cardiac output
Method:
Cardiac output range:
Blood temperature range:
Injectate temperature range:
Waveform display frequency response:
Output parameters:
Cardiac output review:
Catheter sizes:
Injectate volume selection:
Environmental specifications (all modules)
Power requirements:
Power consumption:
Cooling:
Heat dissipation:
Operating temperature:
Operating humidity:
Storage temperature:
Storage humidity:
Physical specifications (Tram modules)
Height:
Width:
Depth:
Weight:
A-14
Selectable saturation 0 to 100% equivalent 0 to 1V
SpO2: 0 to 105%; PPR: 0 to 350 beats per minute.
Thermal dilution
0.2 to 15 liters per minute
30°C to 42°C (86°F to 107°F)
0°C to 30°C (32°F to 86°F)
0 to 10 Hz (-3 dB)
Cardiac output, blood temperature, injectate temperature, real-time cardiac output
waveform
Accept / reject individual measurements and store average
5, 6, 7, 7.5, 8, French
3, 5, 10 cc
16.75 VDC
10 Watts (module alone)
Forced air
Tram 451, 451M, 451N, 851, 851M, 851N: 34 Btu/hr
All others: 31 Btu/hr
10°C to 35°C (50°F to 95°F)
40% to 95% (noncondensing)
-10°C to 50°C (14°F to 122°F)
0% to 95% (noncondensing)
8.1 cm (3.2 in)
11.4 cm (4.5 in)
29.5 cm (11.6 in)
2.1 kg (4.7 lb)
Solar 8000M/i patient monitor
2026265-075C

Hide quick links:

Advertisement

Table of Contents
loading