Recommended Separation Distances - GE Corometrics 250 Series Service Manual

Table of Contents

Advertisement

Recommended Separation Distances

Rated Maximum Output
Power of Transmitter in
Watts
0.01
0.1
1
10
100
a
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
C-6
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
The table below provides the recommended separation distances (in meters)
between portable and mobile RF communications equipment and the 250 Series
Maternal/Fetal Monitor.
The 250 Series Maternal/Fetal Monitor is intended for use in the electromagnetic
environment on which radiated RF disturbances are controlled. The customer or the
user of the 250 Series Maternal/Fetal Monitor can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the 250 Series Maternal/Fetal
Monitor as recommended below, according to the maximum output power of the
communications equipment.
Separation Distance in Meters (m) According to Frequency of Transmitter
a
150 kHz to 80 MHz
P
d = 1.2
0.12
0.38
1.2
3.8
12
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [d] in meters (m) can be estimated using the
equitation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE:
These guidelines may not apply in all instances. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
250 Series Maternal/Fetal Monitor
2020551-001
a
80 MHz to 800 MHz
P
d = 1.2
0.12
0.38
1.2
3.8
12
a
800 MHz to 2.5 GHz
P
d = 2.3
0.23
0.73
2.3
7.3
23
Revision D