Cleaning

Cleaning precautions

Exterior cleaning

Cleaning the display
4-4
Maintenance and checkout: Cleaning
NOTE
See
"Cleaning, disinfecting and storing GE ECG cables and
leadwires"
on page 4-5 for instructions specific to GE ECG cables and
leadwires.
Use one of the following approved solutions:
Cidex solution
Sodium hypochlorite bleach (diluted)
Mild soap (diluted)
Lint-free cloth
Dust remover (compressed air)
To avoid damage to the equipment surfaces, never use the following
cleaning agents:
organic solvents,
ammonia based solutions,
acetone solution,
alcohol based cleaning agents,
Betadine solution,
a wax containing a cleaning substance, or
abrasive cleaning agents.
Clean the exterior surfaces with a clean, lint-free cloth and one of the
cleaning solutions listed in the table above.
Wring the excess solution from the cloth. Do not drip any liquid into
open vents, switches, plugs, or connectors.
Dry the surfaces with a clean cloth or paper towel.
To clean the display, follow the recommendations of the display's
manufacturer. In general you will need to use a soft, clean, lint-free cloth
dampened with a glass cleaner.
CAUTION
To avoid getting liquid into connector openings, do not
spray glass cleaning or general cleaning solutions
directly onto the product's surface.
Solar 8000M/i patient monitor
2026265-075C