Set Line Frequency; Set Defib Sync Voltage And Pulse Width - GE Solar 8000M/i Service Manual

Patient monitor
Table of Contents

Advertisement

Set line frequency

Set defib sync voltage and pulse width

2026265-075C
Configuration: Configuring a patient monitor
Use the Boot Code SERVICE MENU to configure or change the monitor
line frequency to 50 or 60 Hz. The default is 60 Hz.
Activate the Boot Code:
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select 4 SET CONFIGURATION menu option.
5. In the Configuration Menu, select Line Frequency then choose 50
Hz or 60 Hz line frequency.
The Solar 8000M/i patient monitor controls the analog out signal used to
trigger a defibrillator. Refer to the defibrillator manufacturer's manual
for the required pulse amplitude and duration.
Use the Boot Code SERVICE MENU to configure or change the
MARKER OUT signal of the DEFIB SYNC connector on a Solar ECG
module.
Activate the Boot Code:
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select SERVICE MENU.
5. Select 4 SET CONFIGURATION menu option.
6. In the Configuration Menu, select:
1 Defib Sync Voltage: and choose 5V or 12V amplitude.
2 Defib Sync Pulse Width: and choose 10 ms or 100 ms for
pulse duration.
Solar 8000M/i patient monitor
6-7