MONITOR SETTINGS
5-8
Troubleshooting: Service menus
This menu selection provides the following submenus: (See Chapter 6,
"Configuration"
, for detailed procedures.)
SET UNIT NAME
This menu selection allows changes to the care unit name. After initial
setup, this name should not be changed or communication to the central
station will be corrupted. Note that the care unit name must be
registered exactly the same in the central station and the patient
monitor.
SET BED NUMBER
This menu selection allows changes to the bed number. After initial
setup, this number should not be changed or communication to the
central station will be corrupted. Note that the bed number must be
registered exactly the same in the central station and the patient
monitor.
SET LOCATION ID
Enter the location identification for the MUSE system.
SET SITE NUMBER
A site number may be necessary if your institution has multiple
buildings. A MUSE system can be used as a site number.
STORE DEFAULTS FOR NETWORK TRANSFER
This menu selection saves the monitor defaults for transfer to another
monitor with the same software, patient monitor type, and country code.
It stores all monitor defaults, custom default names, and 12 SL location
and site.
SET INTERNET ADDRESS
This menu selection allows changes to the internet (IP) address.
WARNING
Duplication of an internet (IP) address on a network
causes lost data. If you change the factory assigned
internet address, you must first record all other internet
addresses used on your network to avoid duplication.
An incorrect internet address may also prevent the monitor from viewing
other monitors on the network. Whether or not this can occur depends on
the network topology at the installed site.
Solar 8000M/i patient monitor
2026265-075C