GE Solar 8000M/i Service Manual page 116

Patient monitor
Table of Contents

Advertisement

5-24
Troubleshooting: Reviewing error/event logs
For some categories of TRAM-NET network errors, two additional lines
are added to the error log entry.
Name
Network Error
Identifies that a network error occurred.
Channel Number
Identifies the network channel exhibiting the error.
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Levels
Continue
The event or error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log
entries will have a severity of CONTINUE.
Fatal
The event or error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a warm
start.
Forced Restart
The system was restarted by a known condition (internet
address change, video test, etc).
Solar 8000M/i patient monitor
Definition of parameters
Description
Definition of severity levels
Description
2026265-075C