GE Solar 8000M/i Service Manual page 136

Patient monitor
Table of Contents

Advertisement

Copy stored monitor defaults
6-12
Configuration: Advanced user procedures
WARNING
Loss of Defaults. Copying monitor defaults from another
monitor erases all of the monitor defaults on the current
monitor.
To copy monitor defaults from the server monitor, the client monitor
must be in the DISCHARGED state. If the client monitor is set to the
OR patient-monitor type, it must not be in COMBO or ROVER COMBO
admit mode.
NOTE
Make sure that the client monitor shares the same configuration as
the server monitor.
1. At the client monitor, start from the MAIN menu and select MORE
MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the TRIM KNOB control to select the day and
month from monitor screen with leading zeros (e.g., July 4 = 0407).
3. Select COPY UNIT DEFAULTS.
4. Select SELECT UNIT TO COPY MONITOR DEFAULTS FROM.
5. From the list of units on the Unity Network, choose the server
monitor.
6. Select SELECT BED TO COPY MONITOR DEFAULTS FROM.
7. Scroll through the list of beds within the selected unit until the
desired bed is found. Select it, and answer YES to the confirmation
popup menu.
NOTE
After transferring monitor defaults, the first set of defaults is
automatically activated. If another set of defaults is desired, the user
must manually select it from the RECALL DEFAULTS menu.
After copying monitor defaults from another bed (the server monitor),
verify that the defaults were transferred and arrhythmia levels are as
desired.
Solar 8000M/i patient monitor
2026265-075C