Appendix A – Technical specifications: Tram modules and Solar parameter functionality
Input specifications:
Offset:
Input impedance:
Common mode:
Differential:
Output specifications:
Gain:
Frequency response:
Gain stability:
Zero balance range:
Zero balance accuracy:
Zero balance drift:
Common mode rejection:
Noise:
Accuracy:
Displayed frequency response:
Display scale selections:
Analog output:
Alarms:
Noninvasive blood pressure
(Tram 415, 451M, 451N, 851, 851M, 851N, 200SL, 400SL, 800A, 800SL, 250SL, 450SL, 850A, 850SL and 300 only)
Measurement technique:
Displayed parameters:
Systolic pressure range:
Adult:
Pediatric:
Neonatal:
Diastolic pressure range:
Adult:
Pediatric:
Neonatal:
Mean pressure range:
Adult:
Pediatric:
Neonatal:
Cuff pressure range:
Adult:
Pediatric:
Neonatal:
2026265-075C
±150 mmHg
>100 K at 50/60 Hz
>100 K from dc to 60 Hz
976 ±1%
dc to 50 Hz (+0/-3 dB)
<±0.1%/°C, and <±0.1% over any 24 hour period
±150 mmHg
±1 mmHg
±1 mmHg over 24 hours
>60 dB at 60 Hz
<5 mV peak to peak from dc to 30 Hz
±2% or ±1 mmHg, whichever is greater (exclusive of transducer)
0 to 12 Hz or 0 to 40 Hz (-3dB) user-selectable
0-30, 0-40, 0-60, 0-100, 0-160, 0-200, 0-300 mmHg
1 V / 100 mmHg
User selectable upper and lower limits for systolic, diastolic, and mean pressures; range -
99 to 350 mmHg
Oscillometric
Systolic, diastolic, and mean pressures, pulse rate, time of last measurement
30 to 275 mmHg
30 to 240 mmHg
30 to 135 mmHg
10 to 220 mmHg
10 to 200 mmHg
10 to 110 mmHg
20 to 260 mmHg
20 to 235 mmHg
20 to 125 mmHg
0 to 300 mmHg
0 to 250 mmHg
0 to 150 mmHg
Solar 8000M/i patient monitor
A-11