Solar 8000M/i display
Purchaser's responsibility
Medical-grade displays
A-22
Appendix A – Technical specifications: Solar 8000M/i display
The Solar 8000M/i patient monitor is currently available with the
following GE displays:
15-inch, medical-grade, digital flat panel display with touchscreen
15-inch, medical-grade, analog/digital flat panel display with or
without touchscreen
19-inch, medical-grade, analog/digital flat panel display with or
without touchscreen
Off-the-shelf (non-medical grade) displays are also compatible (see the
required and recommended specifications in this appendix).
WARNING
Do not connect a monochrome display to the Solar
8000M/i patient monitor. Visual alarm messages may not
appear properly.
WARNING
Use of non-GE displays with the Solar 8000M/i monitor
may result in loss of visual alarms and patient
monitoring. Display functionality must be verified for
proper operation with the Solar 8000M/i patient monitor
in the clinical environment prior to placing into service.
The display purchaser is responsible for meeting Solar 8000M/i display
specifications. GE does not make recommendations regarding specific
display models other than those it offers for sale. For questions regarding
display specifications or compatibility of displays not purchased from
GE, contact the display manufacturer.
The GE warranty only applies to equipment purchased from GE. Service
repairs resulting from failures of equipment not purchased from GE are
billable.
NOTE
An isolation transformer must be used with a computer-grade
display to meet UL and EN specifications.
The Solar 8000M/i patient monitor with a medical-grade display meets
applicable UL and IEC specifications for a medical electrical system. For
this system, an isolation transformer is not required.
Solar 8000M/i patient monitor