Field replaceable units: Disassembly
d. Install the plug in cover.
Install plug here
2. Remove six screws holding the rear panel. Remove the rear panel.
(The processor board cannot be removed without first removing rear
panel.)
3. Remove cable harness to the power supply.
4. Remove five screws holding the board to the chassis.
5. Disconnect speaker cable.
6. Slide the board toward rear to clear front of chassis.
After replacing the processor pcb
1. Clean or remove the old label and apply the new label from the FRU
kit.
New label
2. Perform a cold start. (See
Visually verify that the "performing cold start" message appears
to be sure the a cold start was actually performed.
3. Restore the above settings.
4. If the patient monitor is not on the Unity Network, be sure to set the
time and date.
WARNING
Patient data loss. Do not change the time and date if the
patient monitor is on a network. Doing so will result in
patient data loss.
If the monitor is on a network, verify that the time and date is
automatically set by the network time master.
Solar 8000M/i patient monitor
"Cold start"
on page 5-2.)
504A
010C
2026265-075C