What error codes mean
2026265-075C
Troubleshooting: Reviewing error/event logs
The error log contains more than just operating system errors. Many
events that have an impact upon the system are also entered into the log.
The 700-series of error codes are really system initiated events. Listed
below are some of the event/error codes you might find useful.
Error code
400-4FF
Network errors were detected.
703
Diagnostic tests were completed.
70B
Internet address was changed. The network address for the monitor
was changed. This should only be done by qualified service
personnel.
70C
Video test was completed. This test should only be performed by
qualified service personnel.
70D
Rear power switch was turned off.
70E
Time was changed from this monitor. This helps determine how the
system-wide time may have been altered.
70F
Date was changed from this monitor. This helps determine how the
system-wide date was altered.
710
Incompatible software was detected. If the main processor software
finds that the software operating on the communication software is
incompatible, it turns off the communication (network) controller and
enters this data into the error log. If the monitor won't "talk" to the
network, looking for this entry in the error log is one part of the
troubleshooting process.
714
Incompatible TRAM module software was detected. If the main
processor software finds the software revision in the TRAM is
incompatible, it turns off communication with the TRAM and enters an
error log. Refer to the software compatibility listing in the SOFTWARE
REVISION window.
SERVICE PATIENT DATA MODULE
NOTE
The monitor may be referred to as a display or scope in the error code
descriptions.
Solar 8000M/i patient monitor