Appendix B – Electromagnetic compatibility: Electromagnetic compatibility (EMC)
Guidance and manufacturer's declaration — electromagnetic
immunity
Immunity test
EN 60601 test level
Electrostatic
± 6 kV contact
Discharge (ESD)
± 8 kV air
EN 61000-4-2
Electrical Fast
± 2 kV for power supply lines
Transient/Burst
EN 61000-4-4
±1 kV for input/output lines
Surge
± 1 kV differential mode
EN 61000-4-5
± 2 kV common mode
<5% U
for 0.5 cycles
Voltage dips, short
<40% U
interruptions and
for 5 cycles
voltage variations
<70% U
on power supply
input lines
for 25 cycles
EN 61000-4-11
<5% U
for 5 s
Power Frequency
3 A/m
(50/60 Hz)
Magnetic Field
EN 61000-4-8
2026265-075C
The Solar 8000M/i is intended for use in the electromagnetic
environment specified below. It is the responsibility of the customer or
user to assure that the Solar 8000M/i is used in such an environment.
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 1 kV differential mode
± 2 kV common mode
(>95% dip in U
)
<5% U
t
t
for 0.5 cycles
(>60% dip in U
)
<40% U
t
t
U
) for 5 cycles
t
(>30% dip in U
)
<70% U
t
t
U
) for 25 cycles
t
(>95% dip in U
)
<5% U
t
t
for 5 s
3 A/m
NOTE
U
is the AC mains voltage prior to application of the test level.
t
Solar 8000M/i patient monitor
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power should be that of a typical commercial
or hospital environment.
Mains power should be that of a typical commercial
or hospital environment.
(>95% dip in U
)
t
t
(>60% dip in
Mains power should be that of a typical commercial
t
or hospital environment. If the user of the equipment
requires continued operation during power mains
(>30% dip in
interruptions, it is recommended that the equipment
t
be powered from an uninterruptible power supply or
a battery.
(>95% dip in U
)
t
t
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
B-3