Table of Contents

Advertisement

AMI Codes-II
Maintenance
6.5.
Process
Calibration of
DIS
Zero
50

Calibration

NOTICE: Perform process calibration for free chlorine / total
residual chlorine only if:
• the sample concentration is close to the desired process
value (stable value)
• you are sure that the reagents are mixed completely and
correctly
• if the difference to the manual measurement is significant.
• Keep in mind the accuracy of your manual measurement.
Use CHEMATEST 25 (or equivalent photometer) to determine the
sample disinfectant concentration. Determine the sample disinfec-
tant value by 3 manual DPD measurements. Calculate the average
value. Compare this value to the value, indicated by the AMI.
Enter process value (mg/l = ppm) under menu
cess DIS
Calibration
Process DIS.
Process pH
Standard pH
Possible error message see
A zero is automatically done before each measurement.
3.1.1
Enter
Press 3 x [Exit]
Calibration Errors, p.
3.1.1, p. 78
for Pro-
Process DIS.
Current Value
x.xx ppm
Slope
Process Value
x.xx ppm
Save
<Enter>
Process DIS.
Current Value
x.xx ppm
Slope
Process Value
x.xx ppm
Save
<Enter>
Process DIS.
Current Value
x.xx ppm
Slope
Calibration Successful
65.
A-96.250.571 / 050517
3.1.1.4
x.xxx
3.1.1.4
x.xxx
3.1.1.5
x.xxx

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.

This answer is automatically generated

Nonhlanhla
March 2, 2025

Please share Data Sheet for the Pherispatic Pump

Subscribe to Our Youtube Channel

Table of Contents