AMI Codes-II
Safety Instructions
1.
General
Target
audience
OM Location
Qualification,
Training
4
AMI Codes-II - Operator's Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
Safety Instructions
The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Operator: Qualified person who uses the equipment
for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
The AMI Operator's Manual shall be kept in proximity of the instru-
ment.
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
A-96.250.571 / 050517
Need help?
Do you have a question about the AMI Codes-II and is the answer not in the manual?
Questions and answers
Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.
If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:
1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.
Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.
This answer is automatically generated