Swann AMI Codes-II Operator's Manual page 69

Table of Contents

Advertisement

AMI Codes-II
Troubleshooting
Error
E001
DIS. Alarm high
E002
DIS. Alarm Low
E003
pH Alarm high
E004
pH Alarm low
E005
DIS. too high
E007
Sample Temp. high
E008
Sample Temp. low
E009
Sample Flow high
E010
Sample Flow low
A-96.250.571 / 050517
Description
Corrective action
– check process
– check programmed value in menu
5.3.1.1.1, p. 86
– check process
– check programmed value in menu
5.3.1.1.25, p. 86
– check process
– correct/calibrate pH sensor, see
Calibration, p. 50
– check programmed value in menu, see
5.3.1.2.1, p. 87
– check process
– correct/calibrate pH sensor, see
Calibration, p. 50
– check programmed value in menu, see
5.3.1.2.21, p. 87
– check process
– check sample temperature
– check programmed value in menu
5.3.1.4.1, p. 88
– check sample temperature
– check programmed value in menu
5.3.1.4.2, p. 88
– check sample input pressure
– readjust sample flow
– check programmed value in menu
5.3.1.3.x, p. 87
– check sample input pressure
– readjust sample flow
– clean instrument, see
protective Filter, p. 53
– check programmed value in menu
5.3.1.3.x, p. 87
Cleaning the
67

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.