Swann AMI Codes-II Operator's Manual page 14

Table of Contents

Advertisement

AMI Codes-II
Product Description
Time
interval of a
measurement
12
The measuring interval can be set between 1 and 12 minutes. The
time sequence of a measurement with a measuring interval of
5 min is shown in the diagram below.
The blue bar represents the sample which flows continuously
through the photometer. A short time before the measurement
starts, a zero point measurement is performed. Then the peristaltic
pump starts and a small portion of the reagents [J] and [K] is
pumped into the mixing chamber. Shortly after, if the mixture is in
the photometer, the sample is measured.
J
K
0
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
J
OXYCON ON-LINE DPD
K
OXYCON ON-LINE Buffer Solution
Zero point measurement
Sample measurement
A-96.250.571 / 050517
m

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.

This answer is automatically generated

Nonhlanhla
March 2, 2025

Please share Data Sheet for the Pherispatic Pump