Fill Or Flush Reagent System; Programming - Swann AMI Codes-II Operator's Manual

Table of Contents

Advertisement

AMI Codes-II
Instrument Setup
4.4.
4.5.

Programming

36

Fill or Flush Reagent System

Fill or flush the reagent tubing:
 upon the initial instrument setup,
 after refilling the reagent canisters,
 before a system shut-down to flush the system with deminer-
alized water until no more reagent is left in the system.
Service
Verification
Fill System
Cleaning
Fill System
Progress
<Enter> to stop
Fill System
Progress
Done
1 Check tubing and flow cell for leaks and repair if necessary.
2 Let the instrument run continuously for 1 hour.
Programming
Program all parameters for external devices (interface, recorders,
etc.)
Program all parameters for instrument operation (disinfectant, lim-
its, alarms).
Program the DPD value of the Verikit in menu <Installation/
Sensors/Ref. Verification>.
If pH option is installed, program the two buffers you want to use for
calibration in menu <Installation/Sensors/Standards>.
See
Program List and Explanations, p.
Navigate to menu< Maintenance /
3.2.2
Service/Fill system>.
Press [Enter].
The peristaltic pump is activated for
3.2.2.5
1.5 minutes.
Press [Exit] 4 x to go back to the oper-
3.2.2.5
ating mode.
75.
A-96.250.571 / 050517

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.

This answer is automatically generated

Nonhlanhla
March 2, 2025

Please share Data Sheet for the Pherispatic Pump

Table of Contents