Reagents For Free Chlorine, Chlorine Dioxide, Bromine And Iodine - Swann AMI Codes-II Operator's Manual

Table of Contents

Advertisement

AMI Codes-II
Maintenance
Canister set up
6.3.1
Prepare
Oxycon
On-line DPD
46
A
B
C
D
2L
2L
E
OXYCON ON-LINE
OXYCON ON-LINE
Pufferlösung
DPD-Reagenz
Tampon
DPD Réactif
Buffer Solution
DPD Reagent
F
G
Reagents for Free Chlorine, Chlorine Dioxide, Bromine
and Iodine
To prepare 2 L of reagents the following chemicals are necessary:
 1 x 50 ml of Oxycon On-line DPD
 1 x 240 gr of Oxycon On-line Buffer
 4 liters of demineralized water
1 Rinse the canister [E] labelled "OXYCON ON LINE DPD-Re-
agent" with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
A
Suction lance without
level detector
(canister 1)
B
Suction lance with
level detector
(canister 2)
C
Level detector
D
2 L mark
E
Canister 1:
Oxycon on-line DPD
F
Canister 2:
Oxycon on-line
Buffer
G
Holder
A-96.250.571 / 050517

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.

This answer is automatically generated

Nonhlanhla
March 2, 2025

Please share Data Sheet for the Pherispatic Pump